Summary
Type: SPE
Application #: 304031
Catalog#: 730014 - SPE Cartridges, Chromabond, C18 ec, 6ml, PP, 500mg, 30/pk
Detection:
Conditions: Column conditioning: 6 ml methanol, then 6 ml water. (Do not let run the column dry.)
Sample aspiration: The prepared sample is passed through the column by vacuum or pressure (max. 1000 ml sample volume)
Washing: With 3 ml methanol - water (5:95, v/v) are washed impurities
Elution: Analytes are eluted with 2 x 2 ml methanol or acetone
Authors: SPE department
Source: Macherey-Nagel